A Prototype Scintillator Real-Time Beam Monitor for Ultra-high Dose Rate Radiotherapy

Fuente: arXiv
Enregistré dans:
Détails bibliographiques
Auteurs principaux: Levin, Daniel S., Friedman, Peter S., Ferretti, Claudio, Ristow, Nicholas, Tecchio, Monica, Litzenberg, Dale W., Bashkirov, Vladimir, Schulte, Reinhard
Format: Preprint
Publié: 2023
Sujets:
Accès en ligne:
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!
_version_ 1866911791767879680
author Levin, Daniel S.
Friedman, Peter S.
Ferretti, Claudio
Ristow, Nicholas
Tecchio, Monica
Litzenberg, Dale W.
Bashkirov, Vladimir
Schulte, Reinhard
author_facet Levin, Daniel S.
Friedman, Peter S.
Ferretti, Claudio
Ristow, Nicholas
Tecchio, Monica
Litzenberg, Dale W.
Bashkirov, Vladimir
Schulte, Reinhard
contents FLASH Radiotherapy (RT) is a potentially new cancer radiotherapy technique where an entire therapeutic dose is delivered in about 0.1 s and at ~1000 times higher dose rate than in conventional RT. For clinical trials to be conducted safely, precise and fast beam monitoring that can generate an out-of-tolerance beam interrupt is required. A FLASH Beam Scintillator Monitor (FBSM) is being developed based in part on a novel proprietary inorganic hybrid scintillator material. The FBSM provides large area coverage, low mass profile, linear response over a broad dynamic range, radiation tolerance, and real-time analysis IEC-compliant fast beam-interrupt signal. This paper includes the design concept and test results from a prototype device in radiation beams that include heavy ions, FLASH level dose per pulse electron beams, anda hospital radiotherapy clinic with electron beams. Results include image quality, response linearity, radiation hardness, spatial resolution, and real-time data processing. The scintillator showed a small -0.02%/kGy signal decrease after a 212 kGy cumulative dose resulting from continuous exposure for 15 minutes at a FLASH compatible dose rate of 234 Gy/s. These tests established the linear response of the FBSM with respect to dose per pulse. Comparison with commercial Gafchromic film indicates that the FBSM produces a high resolution 2D beam image and can reproduce a nearly identical beam profile. At 20 kfps or 50 microsec/frame, the real-time FPGA based computation and analysis of beam position, beam shape, and beam dose takes < 1 microsec.
format Preprint
id arxiv_https___arxiv_org_abs_2305_15306
institution arXiv
publishDate 2023
record_format arxiv
spellingShingle A Prototype Scintillator Real-Time Beam Monitor for Ultra-high Dose Rate Radiotherapy
Levin, Daniel S.
Friedman, Peter S.
Ferretti, Claudio
Ristow, Nicholas
Tecchio, Monica
Litzenberg, Dale W.
Bashkirov, Vladimir
Schulte, Reinhard
Medical Physics
FLASH Radiotherapy (RT) is a potentially new cancer radiotherapy technique where an entire therapeutic dose is delivered in about 0.1 s and at ~1000 times higher dose rate than in conventional RT. For clinical trials to be conducted safely, precise and fast beam monitoring that can generate an out-of-tolerance beam interrupt is required. A FLASH Beam Scintillator Monitor (FBSM) is being developed based in part on a novel proprietary inorganic hybrid scintillator material. The FBSM provides large area coverage, low mass profile, linear response over a broad dynamic range, radiation tolerance, and real-time analysis IEC-compliant fast beam-interrupt signal. This paper includes the design concept and test results from a prototype device in radiation beams that include heavy ions, FLASH level dose per pulse electron beams, anda hospital radiotherapy clinic with electron beams. Results include image quality, response linearity, radiation hardness, spatial resolution, and real-time data processing. The scintillator showed a small -0.02%/kGy signal decrease after a 212 kGy cumulative dose resulting from continuous exposure for 15 minutes at a FLASH compatible dose rate of 234 Gy/s. These tests established the linear response of the FBSM with respect to dose per pulse. Comparison with commercial Gafchromic film indicates that the FBSM produces a high resolution 2D beam image and can reproduce a nearly identical beam profile. At 20 kfps or 50 microsec/frame, the real-time FPGA based computation and analysis of beam position, beam shape, and beam dose takes < 1 microsec.
title A Prototype Scintillator Real-Time Beam Monitor for Ultra-high Dose Rate Radiotherapy
topic Medical Physics
url https://arxiv.org/abs/2305.15306