Noise-induced servo errors in optical clocks utilizing Rabi interrogation

Fuente: arXiv
Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Lindvall, T., Wallin, A. E., Hanhijärvi, K. J., Fordell, T.
Format: Preprint
Veröffentlicht: 2023
Schlagworte:
Online-Zugang:
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
_version_ 1866916088898387968
author Lindvall, T.
Wallin, A. E.
Hanhijärvi, K. J.
Fordell, T.
author_facet Lindvall, T.
Wallin, A. E.
Hanhijärvi, K. J.
Fordell, T.
contents We show that in optical clocks based on Rabi interrogation, both laser-frequency and magnetic-field flicker ($1/f$) noise with zero mean can lead to servo errors at the $10^{-18}$ level if the negative-detuning (red) and positive-detuning (blue) sides of the transition are always probed in the same order. This is due to the strong correlations of flicker noise in combination with an imbalance in the response of the servo discriminator to positive and negative differential frequency noise between the red- and blue-side probing. This imbalance is particularly large for a normalized discriminator. We derive an analytical expression for the servo error based on the correlation function of the laser-frequency or magnetic-field noise and compare it to numerical servo simulations to demonstrate how the error depends on the noise level, servo parameters, and probing sequence. We also show that the servo error can be avoided by normalizing the discriminator with a moving mean or by reversing the red/blue probing order for every second servo cycle.
format Preprint
id arxiv_https___arxiv_org_abs_2306_14781
institution arXiv
publishDate 2023
record_format arxiv
spellingShingle Noise-induced servo errors in optical clocks utilizing Rabi interrogation
Lindvall, T.
Wallin, A. E.
Hanhijärvi, K. J.
Fordell, T.
Optics
Atomic Physics
Instrumentation and Detectors
We show that in optical clocks based on Rabi interrogation, both laser-frequency and magnetic-field flicker ($1/f$) noise with zero mean can lead to servo errors at the $10^{-18}$ level if the negative-detuning (red) and positive-detuning (blue) sides of the transition are always probed in the same order. This is due to the strong correlations of flicker noise in combination with an imbalance in the response of the servo discriminator to positive and negative differential frequency noise between the red- and blue-side probing. This imbalance is particularly large for a normalized discriminator. We derive an analytical expression for the servo error based on the correlation function of the laser-frequency or magnetic-field noise and compare it to numerical servo simulations to demonstrate how the error depends on the noise level, servo parameters, and probing sequence. We also show that the servo error can be avoided by normalizing the discriminator with a moving mean or by reversing the red/blue probing order for every second servo cycle.
title Noise-induced servo errors in optical clocks utilizing Rabi interrogation
topic Optics
Atomic Physics
Instrumentation and Detectors
url https://arxiv.org/abs/2306.14781