Symmetric-Reciprocal-Match Method for Vector Network Analyzer Calibration

Fuente: arXiv
Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Hatab, Ziad, Gadringer, Michael Ernst, Bösch, Wolfgang
Format: Preprint
Veröffentlicht: 2023
Schlagworte:
Online-Zugang:
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
_version_ 1866929269123317760
author Hatab, Ziad
Gadringer, Michael Ernst
Bösch, Wolfgang
author_facet Hatab, Ziad
Gadringer, Michael Ernst
Bösch, Wolfgang
contents This paper proposes a new approach, the symmetric-reciprocal-match (SRM) method, for calibrating vector network analyzers (VNAs). The method involves using multiple symmetric one-port loads, a two-port reciprocal device, and a matched load. The load standards consist of two-port symmetric one-port devices, and at least three unique loads are used. However, the specific impedances of the loads are not specified. The reciprocal device can be any transmissive device, although a non-reciprocal device can also be used if only the one-port error boxes are of interest. The matched load is fully defined and can be asymmetric. We numerically demonstrated the proposed method's accuracy with synthetic data and with measurements of coaxial standards using a commercial short-open-load-reciprocal (SOLR) calibration kit with verification standards. An advantage of the proposed method is that only the match standard is defined, whereas the remaining standards are partially defined, either through symmetry or reciprocity.
format Preprint
id arxiv_https___arxiv_org_abs_2309_02886
institution arXiv
publishDate 2023
record_format arxiv
spellingShingle Symmetric-Reciprocal-Match Method for Vector Network Analyzer Calibration
Hatab, Ziad
Gadringer, Michael Ernst
Bösch, Wolfgang
Signal Processing
Instrumentation and Detectors
This paper proposes a new approach, the symmetric-reciprocal-match (SRM) method, for calibrating vector network analyzers (VNAs). The method involves using multiple symmetric one-port loads, a two-port reciprocal device, and a matched load. The load standards consist of two-port symmetric one-port devices, and at least three unique loads are used. However, the specific impedances of the loads are not specified. The reciprocal device can be any transmissive device, although a non-reciprocal device can also be used if only the one-port error boxes are of interest. The matched load is fully defined and can be asymmetric. We numerically demonstrated the proposed method's accuracy with synthetic data and with measurements of coaxial standards using a commercial short-open-load-reciprocal (SOLR) calibration kit with verification standards. An advantage of the proposed method is that only the match standard is defined, whereas the remaining standards are partially defined, either through symmetry or reciprocity.
title Symmetric-Reciprocal-Match Method for Vector Network Analyzer Calibration
topic Signal Processing
Instrumentation and Detectors
url https://arxiv.org/abs/2309.02886