contents We have established a method to measure picograms-per-gram (pg g$^{-1}$) levels of $^{238}$U and $^{232}$Th in scintillation films by combining the dry ashing method and inductively coupled plasma mass spectrometry. Trace amounts of $^{238}$U and $^{232}$Th were measured in up to 2~g of the scintillation film with almost 100% collection efficiency. This paper details the experimental procedure, including the pretreatment of the samples and labware, detection limit of the method, collection efficiencies of $^{238}$U and $^{232}$Th, and measurement of $^{238}$U and $^{232}$Th in a polyethylene naphthalate film. This method is also applicable to $^{238}$U and $^{232}$Th measurements in other low-background organic materials for rare event search experiments.
format Preprint
id arxiv_https___arxiv_org_abs_2405_04901
institution arXiv
publishDate 2024
record_format arxiv
spellingShingle Development of a method to measure trace level of uranium and thorium in scintillation films
Ichimura, K.
Chiba, K.
Gando, Y.
Ikeda, H.
Kishimoto, Y.
Kurasawa, M.
Nemoto, K.
Sakaguchi, A.
Takaku, Y.
Sakakieda, Y.
Instrumentation and Detectors
We have established a method to measure picograms-per-gram (pg g$^{-1}$) levels of $^{238}$U and $^{232}$Th in scintillation films by combining the dry ashing method and inductively coupled plasma mass spectrometry. Trace amounts of $^{238}$U and $^{232}$Th were measured in up to 2~g of the scintillation film with almost 100% collection efficiency. This paper details the experimental procedure, including the pretreatment of the samples and labware, detection limit of the method, collection efficiencies of $^{238}$U and $^{232}$Th, and measurement of $^{238}$U and $^{232}$Th in a polyethylene naphthalate film. This method is also applicable to $^{238}$U and $^{232}$Th measurements in other low-background organic materials for rare event search experiments.
title Development of a method to measure trace level of uranium and thorium in scintillation films
topic Instrumentation and Detectors