Auteurs principaux: Li, Yuanxia, Zhao, Jie, Ding, Yayun, Hu, Tao, Ye, Jiaxuan, Fang, Jian, Wen, Liangjian
Format: Preprint
Publié: 2024
Sujets:
Accès en ligne:https://arxiv.org/abs/2405.06326
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!
Table des matières:
  • Liquid scintillator (LS) is commonly utilized in experiments seeking rare events due to its high light yield, transparency, and radiopurity. The concentration of $^{238}$U and $^{232}$Th in LS consistently remains below 1 ppq (10$^{-15}$ g/g), and the current screening result is based on a minimum 20-ton detector. Inductively coupled plasma mass (ICP-MS) spectroscopy is well-regarded for its high sensitivity to trace $^{238}$U and $^{232}$Th. This study outlines a method for detecting $^{238}$U and $^{232}$Th in LS at the sub-ppq level using ICP-MS, involving the enrichment of $^{238}$U/$^{232}$Th from the LS through acid extraction. With meticulous cleanliness control, $^{238}$U/$^{232}$Th in approximately 2 kg of LS is concentrated by acid extraction with 0.4 (0.3) pg $^{238}$U ($^{232}$Th) contamination. Three standard adding methods are employed to assess recovery efficiency, including radon daughter, 2,5-diphenyloxazole (PPO), and natural non-existent $^{233}$U/$^{229}$Th. The method detection limit at a 99% confidence level of this approach can reach approximately 0.2-0.3 ppq for $^{238}$U/$^{232}$Th with nearly 100% recovery efficiency.