Collimator-less SPECT System Design for Dynamic Whole-body Imaging

Fuente: arXiv
Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Feng, Yuemeng, Sitek, Arkadiusz, Esfahani, Shadi Abdar, Sabet, Hamid
Format: Preprint
Veröffentlicht: 2024
Schlagworte:
Online-Zugang:
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
_version_ 1866913902132985856
author Feng, Yuemeng
Sitek, Arkadiusz
Esfahani, Shadi Abdar
Sabet, Hamid
author_facet Feng, Yuemeng
Sitek, Arkadiusz
Esfahani, Shadi Abdar
Sabet, Hamid
contents In this study, we introduce a Compton SPECT system for whole-body imaging of Actinium-225 (225Ac), one of the trending radionuclides for targeted alpha therapy (TAT). The Compton SPECT system enables multi-energy gamma photon detection with higher efficiency compared to mechanically collimated SPECT. The system consists of two detectors, providing a field of view (FOV) adequate for whole-body imaging, while achieving high sensitivity and clinically usable imaging resolution within a reasonable scanning time. This work focuses on the system design and evaluation using the Monte Carlo simulation toolkit Gate. The imaging performance is evaluated at two energy peaks (218 keV, 440 keV), representing the major detectable gamma energies generated from 225Ac. We explore the possibility of using the Compton SPECT system for treatment response monitoring in TAT. Results demonstrate an image resolution of 1.0 cm using a NEMA IQ phantom with 5.7 MBq of 225Ac simulated in a cold background. An image resolution of 1.3 cm can be achieved with a hot to background ratio of 30:1, and a resolution of 3.7 cm can be achieved with an activity ratio of 12:1. The best achievable sensitivity at 10 cm distance to the detector is 0.5% with the two energy windows (211-225 keV, 430-450 keV) selected. The proposed system may serve as an alternative imaging tool for TAT scanning in clinical settings.
format Preprint
id arxiv_https___arxiv_org_abs_2406_05220
institution arXiv
publishDate 2024
record_format arxiv
spellingShingle Collimator-less SPECT System Design for Dynamic Whole-body Imaging
Feng, Yuemeng
Sitek, Arkadiusz
Esfahani, Shadi Abdar
Sabet, Hamid
Medical Physics
In this study, we introduce a Compton SPECT system for whole-body imaging of Actinium-225 (225Ac), one of the trending radionuclides for targeted alpha therapy (TAT). The Compton SPECT system enables multi-energy gamma photon detection with higher efficiency compared to mechanically collimated SPECT. The system consists of two detectors, providing a field of view (FOV) adequate for whole-body imaging, while achieving high sensitivity and clinically usable imaging resolution within a reasonable scanning time. This work focuses on the system design and evaluation using the Monte Carlo simulation toolkit Gate. The imaging performance is evaluated at two energy peaks (218 keV, 440 keV), representing the major detectable gamma energies generated from 225Ac. We explore the possibility of using the Compton SPECT system for treatment response monitoring in TAT. Results demonstrate an image resolution of 1.0 cm using a NEMA IQ phantom with 5.7 MBq of 225Ac simulated in a cold background. An image resolution of 1.3 cm can be achieved with a hot to background ratio of 30:1, and a resolution of 3.7 cm can be achieved with an activity ratio of 12:1. The best achievable sensitivity at 10 cm distance to the detector is 0.5% with the two energy windows (211-225 keV, 430-450 keV) selected. The proposed system may serve as an alternative imaging tool for TAT scanning in clinical settings.
title Collimator-less SPECT System Design for Dynamic Whole-body Imaging
topic Medical Physics
url https://arxiv.org/abs/2406.05220