Burn-in Test and Thermal Performance Evaluation of Silicon Photomultipliers for the JUNO-TAO Experiment

Fuente: arXiv
Saved in:
Bibliographic Details
Main Authors: Chen, X., Cao, G. F., Qu, M. H., Wang, H. W., Anfimov, N., Rybnikov, A., Xu, J. Y., Su, A. Q., Chen, Z. L., Cao, J., Li, Y. C., Qi, M.
Format: Preprint
Published: 2024
Subjects:
Online Access:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1866909265123344384
author Chen, X.
Cao, G. F.
Qu, M. H.
Wang, H. W.
Anfimov, N.
Rybnikov, A.
Xu, J. Y.
Su, A. Q.
Chen, Z. L.
Cao, J.
Li, Y. C.
Qi, M.
author_facet Chen, X.
Cao, G. F.
Qu, M. H.
Wang, H. W.
Anfimov, N.
Rybnikov, A.
Xu, J. Y.
Su, A. Q.
Chen, Z. L.
Cao, J.
Li, Y. C.
Qi, M.
contents This study evaluates more than 4,000 tiles made of Hamamatsu visual-sensitive silicon photomultipier (SiPM), each with dimensions of 5 $\times$ 5 cm$^2$, intended for the central detector of the Taishan Anti-neutrino Observatory (TAO), a satellite experiment of the Jiangmen Underground Neutrino Observatory (JUNO) aimed at measuring the reactor anti-neutrino energy spectrum with unprecedented energy resolution. All SiPM tiles underwent a room temperature burn-in test in the dark for two weeks, while cryogenic testing analyzed the thermal dependence of parameters for some sampled SiPMs. Results from these comprehensive tests provide crucial insights into the long-term performance and stability of the 10 square meters of SiPMs operating at -50°C to detect scintillation photons in the TAO detector. Despite some anomalies awaiting further inspection, all SiPMs successfully passed the burn-in test over two weeks at room temperature, which is equivalent to 6.7 years at -50°C. Results are also used to guide optimal SiPM selection, configuration, and operation, ensuring reliability and sustainability in reactor neutrino measurements. This work also provides insights for a rapid and robust quality assessment in future experiments that employ large-scale SiPMs as detection systems.
format Preprint
id arxiv_https___arxiv_org_abs_2406_12912
institution arXiv
publishDate 2024
record_format arxiv
spellingShingle Burn-in Test and Thermal Performance Evaluation of Silicon Photomultipliers for the JUNO-TAO Experiment
Chen, X.
Cao, G. F.
Qu, M. H.
Wang, H. W.
Anfimov, N.
Rybnikov, A.
Xu, J. Y.
Su, A. Q.
Chen, Z. L.
Cao, J.
Li, Y. C.
Qi, M.
Instrumentation and Detectors
High Energy Physics - Experiment
This study evaluates more than 4,000 tiles made of Hamamatsu visual-sensitive silicon photomultipier (SiPM), each with dimensions of 5 $\times$ 5 cm$^2$, intended for the central detector of the Taishan Anti-neutrino Observatory (TAO), a satellite experiment of the Jiangmen Underground Neutrino Observatory (JUNO) aimed at measuring the reactor anti-neutrino energy spectrum with unprecedented energy resolution. All SiPM tiles underwent a room temperature burn-in test in the dark for two weeks, while cryogenic testing analyzed the thermal dependence of parameters for some sampled SiPMs. Results from these comprehensive tests provide crucial insights into the long-term performance and stability of the 10 square meters of SiPMs operating at -50°C to detect scintillation photons in the TAO detector. Despite some anomalies awaiting further inspection, all SiPMs successfully passed the burn-in test over two weeks at room temperature, which is equivalent to 6.7 years at -50°C. Results are also used to guide optimal SiPM selection, configuration, and operation, ensuring reliability and sustainability in reactor neutrino measurements. This work also provides insights for a rapid and robust quality assessment in future experiments that employ large-scale SiPMs as detection systems.
title Burn-in Test and Thermal Performance Evaluation of Silicon Photomultipliers for the JUNO-TAO Experiment
topic Instrumentation and Detectors
High Energy Physics - Experiment