Optically-Validated Microvascular Phantom for Super-Resolution Ultrasound Imaging

Fuente: arXiv
Saved in:
Bibliographic Details
Main Authors: Raad, Jaime Parra, Lock, Daniel, Liu, Yi-Yi, Solomon, Mark, Peralta, Laura, Christensen-Jeffries, Kirsten
Format: Preprint
Published: 2024
Subjects:
Online Access:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1866929551993470976
author Raad, Jaime Parra
Lock, Daniel
Liu, Yi-Yi
Solomon, Mark
Peralta, Laura
Christensen-Jeffries, Kirsten
author_facet Raad, Jaime Parra
Lock, Daniel
Liu, Yi-Yi
Solomon, Mark
Peralta, Laura
Christensen-Jeffries, Kirsten
contents Super-resolution ultrasound (SRUS) visualises microvasculature beyond the ultrasound diffraction limit (wavelength($λ$)/2) by localising and tracking spatially isolated microbubble contrast agents. SRUS phantoms typically consist of simple tube structures, where diameter channels below 100 $μ$m are not available. Furthermore, these phantoms are generally fragile and unstable, have limited ground truth validation, and their simple structure limits the evaluation of SRUS algorithms. To aid SRUS development, robust and durable phantoms with known and physiologically relevant microvasculature are needed for repeatable SRUS testing. This work proposes a method to fabricate durable microvascular phantoms that allow optical gauging for SRUS validation. The methodology used a microvasculature negative print embedded in a Polydimethylsiloxane to fabricate a microvascular phantom. Branching microvascular phantoms with variable microvascular density were demonstrated with optically validated vessel diameters down to $\sim$ 60 $μ$m ($λ$/5.8; $λ$ =$\sim$ 350 $μ$m). SRUS imaging was performed and validated with optical measurements. The average SRUS error was 15.61 $μ$m ($λ$/22) with a standard deviation error of 11.44 $μ$m. The average error decreased to 7.93 $μ$m ($λ$/44) once the number of localised microbubbles surpassed 1000 per estimated diameter. In addition, the less than 10$\%$ variance of acoustic and optical properties and the mechanical toughness of the phantoms measured a year after fabrication demonstrated their long-term durability. This work presents a method to fabricate durable and optically validated complex microvascular phantoms which can be used to quantify SRUS performance and facilitate its further development.
format Preprint
id arxiv_https___arxiv_org_abs_2409_09031
institution arXiv
publishDate 2024
record_format arxiv
spellingShingle Optically-Validated Microvascular Phantom for Super-Resolution Ultrasound Imaging
Raad, Jaime Parra
Lock, Daniel
Liu, Yi-Yi
Solomon, Mark
Peralta, Laura
Christensen-Jeffries, Kirsten
Medical Physics
Super-resolution ultrasound (SRUS) visualises microvasculature beyond the ultrasound diffraction limit (wavelength($λ$)/2) by localising and tracking spatially isolated microbubble contrast agents. SRUS phantoms typically consist of simple tube structures, where diameter channels below 100 $μ$m are not available. Furthermore, these phantoms are generally fragile and unstable, have limited ground truth validation, and their simple structure limits the evaluation of SRUS algorithms. To aid SRUS development, robust and durable phantoms with known and physiologically relevant microvasculature are needed for repeatable SRUS testing. This work proposes a method to fabricate durable microvascular phantoms that allow optical gauging for SRUS validation. The methodology used a microvasculature negative print embedded in a Polydimethylsiloxane to fabricate a microvascular phantom. Branching microvascular phantoms with variable microvascular density were demonstrated with optically validated vessel diameters down to $\sim$ 60 $μ$m ($λ$/5.8; $λ$ =$\sim$ 350 $μ$m). SRUS imaging was performed and validated with optical measurements. The average SRUS error was 15.61 $μ$m ($λ$/22) with a standard deviation error of 11.44 $μ$m. The average error decreased to 7.93 $μ$m ($λ$/44) once the number of localised microbubbles surpassed 1000 per estimated diameter. In addition, the less than 10$\%$ variance of acoustic and optical properties and the mechanical toughness of the phantoms measured a year after fabrication demonstrated their long-term durability. This work presents a method to fabricate durable and optically validated complex microvascular phantoms which can be used to quantify SRUS performance and facilitate its further development.
title Optically-Validated Microvascular Phantom for Super-Resolution Ultrasound Imaging
topic Medical Physics
url https://arxiv.org/abs/2409.09031