Field output correction factors using a fully characterized plastic scintillation detector

Fuente: arXiv
Guardado en:
Detalles Bibliográficos
Autores principales: Gingras, Luc, Espinosa, Yunuen Cervantes, Beaulieu, Frederic, Besnier, Magali, Coté, Benjamin, Lambert-Girard, Simon, Leblanc, Danahé, Lechasseur, Yoan, Therriault-Proulx, François, Beaulieu, Luc, Archambault, Louis
Formato: Preprint
Publicado: 2024
Materias:
Acceso en línea:
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
_version_ 1866914962390122496
author Gingras, Luc
Espinosa, Yunuen Cervantes
Beaulieu, Frederic
Besnier, Magali
Coté, Benjamin
Lambert-Girard, Simon
Leblanc, Danahé
Lechasseur, Yoan
Therriault-Proulx, François
Beaulieu, Luc
Archambault, Louis
author_facet Gingras, Luc
Espinosa, Yunuen Cervantes
Beaulieu, Frederic
Besnier, Magali
Coté, Benjamin
Lambert-Girard, Simon
Leblanc, Danahé
Lechasseur, Yoan
Therriault-Proulx, François
Beaulieu, Luc
Archambault, Louis
contents As small radiation fields play an ever-increasing role in radiation therapy, accurate dosimetry of these fields becomes critical to ensure high quality dose calculation and treatment optimization. The PRB-0002, a new plastic scintillation detector (PSD) part of the Hyperscint RP-200 dosimetric platform from Medscint, can potentially facilitate small field measurements. In this work our objective is twofold: first we perform a detailed characterization of the field output correction factors of the PRB-0002 over a wide range of field sizes; second we demonstrate how this PSD can be used to determine the field output correction factors for other small field detectors. Uncertainties were carefully considered and a detailed uncertainty budget is provided. EGSnrc Monte Carlo simulations were conducted to determine the impact of detector composition, surrounding materials, dose averaging within the sensitive volume as well as ionization quenching. From these simulations, the field output correction factors of this PSD were determined. Then, by experimental comparisons, field output correction factors for 2 solid state detectors and 3 small volume ion chambers were. With PRB-0002, the material composition was found to be balanced with the ionization quenching making the field output correction factor near unity. For fields between 0.6 $\times$ 0.6 and 30 $\times$ 30~cm$^2$, the field output correction factors of the PRB-0002 were between 1.002 and 0.999 with a total uncertainty of 0.5 %. Analysis of the uncertainty budget showed that an overall uncertainty of 0.59% can be achieved for a 1 $\times$ 1~cm$^2$ using PRB-0002. We conclude that PRB-0002 is a near-ideal detector for small field dosimetry. Furthermore, it can be used to experimentally determine the field output correction factors of other dosimeters with great accuracy.
format Preprint
id arxiv_https___arxiv_org_abs_2410_00999
institution arXiv
publishDate 2024
record_format arxiv
spellingShingle Field output correction factors using a fully characterized plastic scintillation detector
Gingras, Luc
Espinosa, Yunuen Cervantes
Beaulieu, Frederic
Besnier, Magali
Coté, Benjamin
Lambert-Girard, Simon
Leblanc, Danahé
Lechasseur, Yoan
Therriault-Proulx, François
Beaulieu, Luc
Archambault, Louis
Instrumentation and Detectors
Medical Physics
As small radiation fields play an ever-increasing role in radiation therapy, accurate dosimetry of these fields becomes critical to ensure high quality dose calculation and treatment optimization. The PRB-0002, a new plastic scintillation detector (PSD) part of the Hyperscint RP-200 dosimetric platform from Medscint, can potentially facilitate small field measurements. In this work our objective is twofold: first we perform a detailed characterization of the field output correction factors of the PRB-0002 over a wide range of field sizes; second we demonstrate how this PSD can be used to determine the field output correction factors for other small field detectors. Uncertainties were carefully considered and a detailed uncertainty budget is provided. EGSnrc Monte Carlo simulations were conducted to determine the impact of detector composition, surrounding materials, dose averaging within the sensitive volume as well as ionization quenching. From these simulations, the field output correction factors of this PSD were determined. Then, by experimental comparisons, field output correction factors for 2 solid state detectors and 3 small volume ion chambers were. With PRB-0002, the material composition was found to be balanced with the ionization quenching making the field output correction factor near unity. For fields between 0.6 $\times$ 0.6 and 30 $\times$ 30~cm$^2$, the field output correction factors of the PRB-0002 were between 1.002 and 0.999 with a total uncertainty of 0.5 %. Analysis of the uncertainty budget showed that an overall uncertainty of 0.59% can be achieved for a 1 $\times$ 1~cm$^2$ using PRB-0002. We conclude that PRB-0002 is a near-ideal detector for small field dosimetry. Furthermore, it can be used to experimentally determine the field output correction factors of other dosimeters with great accuracy.