Efficient noble gas purification using hot getters and gas circulation by convection

Fuente: arXiv
Saved in:
Bibliographic Details
Main Authors: Teixeira, J. M. R., Henriques, C. A. O., Silva, P. A. O. C., Mano, R. D. P., Santos, J. M. F. Dos, Monteiro, C. M. B.
Format: Preprint
Published: 2024
Subjects:
Online Access:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1866914997165096960
author Teixeira, J. M. R.
Henriques, C. A. O.
Silva, P. A. O. C.
Mano, R. D. P.
Santos, J. M. F. Dos
Monteiro, C. M. B.
author_facet Teixeira, J. M. R.
Henriques, C. A. O.
Silva, P. A. O. C.
Mano, R. D. P.
Santos, J. M. F. Dos
Monteiro, C. M. B.
contents Noble gas radiation detectors with optical readout are gaining popularity in fields like astrophysics and particle physics due to their ability to produce both ionization and scintillation signals in response to ionizing radiation interaction. In addition, the amplification of primary ionization signals can be achieved by promoting secondary scintillation in the gas. Noble gas purity, especially concerning impurities like H$_{2}$O, N$_{2}$, O$_{2}$, CO$_{2}$, and hydrocarbons, greatly influences its performance. These impurities can cause the loss of primary electrons and quench the scintillation signal. A very high purity level of the gas is required. In the early 90's, a simple method was developed for noble gas purification in sealed, small volume (up to few litters) gas radiation detectors. Gas purification is achieved promoting gas circulation through Zr-based hot getters, simply maintaining the gas circulation by convection. The effectiveness of this method has been only confirmed by the energy resolutions achieved in those detectors, which were similar to that achieved in other high-performance noble gas detectors. In this work, we used waveform analysis of the primary and secondary scintillation signals and we were able to evaluate the impact of the attachment and quenching caused by impurities in one of our detectors filled with pure Xe, and estimate upper values for the impurity content in the gas. The maximum overall impurity concentration was estimated to be below 4 ppm, considering nearly all the impurities, and below 82 ppm if N$_{2}$ is considered. The electron lifetime was measured to be 2.1 $\pm$ 0.1 ms, in line with those achieved in other high-performance optical detectors.
format Preprint
id arxiv_https___arxiv_org_abs_2410_22483
institution arXiv
publishDate 2024
record_format arxiv
spellingShingle Efficient noble gas purification using hot getters and gas circulation by convection
Teixeira, J. M. R.
Henriques, C. A. O.
Silva, P. A. O. C.
Mano, R. D. P.
Santos, J. M. F. Dos
Monteiro, C. M. B.
Instrumentation and Detectors
Noble gas radiation detectors with optical readout are gaining popularity in fields like astrophysics and particle physics due to their ability to produce both ionization and scintillation signals in response to ionizing radiation interaction. In addition, the amplification of primary ionization signals can be achieved by promoting secondary scintillation in the gas. Noble gas purity, especially concerning impurities like H$_{2}$O, N$_{2}$, O$_{2}$, CO$_{2}$, and hydrocarbons, greatly influences its performance. These impurities can cause the loss of primary electrons and quench the scintillation signal. A very high purity level of the gas is required. In the early 90's, a simple method was developed for noble gas purification in sealed, small volume (up to few litters) gas radiation detectors. Gas purification is achieved promoting gas circulation through Zr-based hot getters, simply maintaining the gas circulation by convection. The effectiveness of this method has been only confirmed by the energy resolutions achieved in those detectors, which were similar to that achieved in other high-performance noble gas detectors. In this work, we used waveform analysis of the primary and secondary scintillation signals and we were able to evaluate the impact of the attachment and quenching caused by impurities in one of our detectors filled with pure Xe, and estimate upper values for the impurity content in the gas. The maximum overall impurity concentration was estimated to be below 4 ppm, considering nearly all the impurities, and below 82 ppm if N$_{2}$ is considered. The electron lifetime was measured to be 2.1 $\pm$ 0.1 ms, in line with those achieved in other high-performance optical detectors.
title Efficient noble gas purification using hot getters and gas circulation by convection
topic Instrumentation and Detectors
url https://arxiv.org/abs/2410.22483