contents An automated temperature-controlled electrical DC voltage and DC resistance multiple reference standard (MRS) has been developed by Measurements International (MI) with the scientific support from the Istituto Nazionale di Ricerca Metrologica (INRIM). The MRS includes a 10 V, a 1 Ω, and a 10 kΩ standards selectable via a switch unit. This setup allows the artifact calibration of high-end calibrators and multimeters used in low-frequency electrical measurements. The two resistors are high-stability standards from MI, while the 10 V standard is based on a low-noise circuit developed by INRIM in collaboration with MI. A key innovation is the internal real-time clock calendar, which displays the calibration values of the MRS standards and their updated values internally calculated. This ensures reliable use of the MRS standards over extended periods between calibrations, effectively minimizing uncertainties due to their drift. The standards are housed in a thermal box, minimizing temperature variations. The MRS standards meet the uncertainty requirements defined by calibrators and multimeters manufacturers for artifact calibration and can also serve as laboratory references or travelling standards for interlaboratory comparisons (ILCs). MI is currently commercializing the MRS.
format Preprint
id arxiv_https___arxiv_org_abs_2501_10697
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle Automated DC Voltage and DC Resistance Real-time Multiple Standard for Artifact Calibration of Calibrators and multimeters
Gallianaa, F.
Capra, P.
Cerri, R.
Francese, C.
Tet, M. Lanzillotti L. Roncaglione
Pennecchi, F.
Pollarolo, A.
Instrumentation and Detectors
An automated temperature-controlled electrical DC voltage and DC resistance multiple reference standard (MRS) has been developed by Measurements International (MI) with the scientific support from the Istituto Nazionale di Ricerca Metrologica (INRIM). The MRS includes a 10 V, a 1 Ω, and a 10 kΩ standards selectable via a switch unit. This setup allows the artifact calibration of high-end calibrators and multimeters used in low-frequency electrical measurements. The two resistors are high-stability standards from MI, while the 10 V standard is based on a low-noise circuit developed by INRIM in collaboration with MI. A key innovation is the internal real-time clock calendar, which displays the calibration values of the MRS standards and their updated values internally calculated. This ensures reliable use of the MRS standards over extended periods between calibrations, effectively minimizing uncertainties due to their drift. The standards are housed in a thermal box, minimizing temperature variations. The MRS standards meet the uncertainty requirements defined by calibrators and multimeters manufacturers for artifact calibration and can also serve as laboratory references or travelling standards for interlaboratory comparisons (ILCs). MI is currently commercializing the MRS.