Shoot-through layers in upright proton arcs unlock advantages in plan quality and range verification

Fuente: arXiv
Saved in:
Bibliographic Details
Main Authors: Engwall, Erik, Mikhalev, Victor, Sundström, Johan, Marthin, Otte, Wase, Viktor
Format: Preprint
Published: 2025
Subjects:
Online Access:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1866916910382186496
author Engwall, Erik
Mikhalev, Victor
Sundström, Johan
Marthin, Otte
Wase, Viktor
author_facet Engwall, Erik
Mikhalev, Victor
Sundström, Johan
Marthin, Otte
Wase, Viktor
contents Background: Upright proton therapy with compact delivery systems has the potential to reduce costs for treatments but could also lead to broadening of the beam penumbra due to energy selection close to the patient. Purpose: This study aims at combining upright static proton arcs with additional layers of shoot-through (ST) protons to sharpen the beam penumbra and improve plan quality for such systems. Methods: We examined various treatment plans for a virtual phantom: 3-beam IMPT, static arc (Arc) with/without ST (Arc+ST), and with/without collimation (+Coll). In the virtual phantom three different targets were utilized to study the effect on conformity index (CI), homogeneity index (HI), robustness and mean dose to the phantom volume. The phantom study was complemented with a head-and-neck (H&N) patient case with a similar set of plans. The delivery time for all plans was estimated. A range verification concept that determines residual ranges of the ST protons was studied in simulated scenarios for the H&N case. Results: In the phantom study, the Arc+ST plans show superior CI, HI and target robustness compared to the Arc+Coll plans. For the Arc plans without ST, the collimated plans perform better than the uncollimated plans. On the other hand, for Arc+ST, collimation has little impact on CI, HI and robustness. For the H&N case, similar improvements for Arc+ST can be seen. These results imply that no aperture is needed when combining arcs with ST, which in turn substantially reduces treatment times. The range verification simulation shows that the method is sensitive to detect SPR errors, setup errors and changes in the patient anatomy. Conclusions: Combining proton arcs and ST layers can enhance compact upright proton solutions by improving plan quality. It is also tailored for the inclusion of a fast and straightforward residual range verification method.
format Preprint
id arxiv_https___arxiv_org_abs_2503_12904
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle Shoot-through layers in upright proton arcs unlock advantages in plan quality and range verification
Engwall, Erik
Mikhalev, Victor
Sundström, Johan
Marthin, Otte
Wase, Viktor
Medical Physics
Background: Upright proton therapy with compact delivery systems has the potential to reduce costs for treatments but could also lead to broadening of the beam penumbra due to energy selection close to the patient. Purpose: This study aims at combining upright static proton arcs with additional layers of shoot-through (ST) protons to sharpen the beam penumbra and improve plan quality for such systems. Methods: We examined various treatment plans for a virtual phantom: 3-beam IMPT, static arc (Arc) with/without ST (Arc+ST), and with/without collimation (+Coll). In the virtual phantom three different targets were utilized to study the effect on conformity index (CI), homogeneity index (HI), robustness and mean dose to the phantom volume. The phantom study was complemented with a head-and-neck (H&N) patient case with a similar set of plans. The delivery time for all plans was estimated. A range verification concept that determines residual ranges of the ST protons was studied in simulated scenarios for the H&N case. Results: In the phantom study, the Arc+ST plans show superior CI, HI and target robustness compared to the Arc+Coll plans. For the Arc plans without ST, the collimated plans perform better than the uncollimated plans. On the other hand, for Arc+ST, collimation has little impact on CI, HI and robustness. For the H&N case, similar improvements for Arc+ST can be seen. These results imply that no aperture is needed when combining arcs with ST, which in turn substantially reduces treatment times. The range verification simulation shows that the method is sensitive to detect SPR errors, setup errors and changes in the patient anatomy. Conclusions: Combining proton arcs and ST layers can enhance compact upright proton solutions by improving plan quality. It is also tailored for the inclusion of a fast and straightforward residual range verification method.
title Shoot-through layers in upright proton arcs unlock advantages in plan quality and range verification
topic Medical Physics
url https://arxiv.org/abs/2503.12904