High-energy X-ray phase-contrast CT of an adult human chest phantom

Fuente: arXiv
Saved in:
Bibliographic Details
Main Authors: Ahlers, Jannis N., D'Amico, Lorenzo, Bast, Henriette, Costello, Lucy F., Donnelley, Martin, Alloo, Samantha J., Harker, Stephanie A., How, Ying Ying, Croughan, Michelle K., Pollock, James A., Hausermann, Daniel, Maksimenko, Anton, Hall, Christopher, Gureyev, Timur E., Nesterets, Yakov I., Kitchen, Marcus J., Pavlov, Konstantin M., Morgan, Kaye S.
Format: Preprint
Published: 2025
Subjects:
Online Access:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1866912640011337728
author Ahlers, Jannis N.
D'Amico, Lorenzo
Bast, Henriette
Costello, Lucy F.
Donnelley, Martin
Alloo, Samantha J.
Harker, Stephanie A.
How, Ying Ying
Croughan, Michelle K.
Pollock, James A.
Hausermann, Daniel
Maksimenko, Anton
Hall, Christopher
Gureyev, Timur E.
Nesterets, Yakov I.
Kitchen, Marcus J.
Pavlov, Konstantin M.
Morgan, Kaye S.
author_facet Ahlers, Jannis N.
D'Amico, Lorenzo
Bast, Henriette
Costello, Lucy F.
Donnelley, Martin
Alloo, Samantha J.
Harker, Stephanie A.
How, Ying Ying
Croughan, Michelle K.
Pollock, James A.
Hausermann, Daniel
Maksimenko, Anton
Hall, Christopher
Gureyev, Timur E.
Nesterets, Yakov I.
Kitchen, Marcus J.
Pavlov, Konstantin M.
Morgan, Kaye S.
contents Propagation-based phase-contrast X-ray imaging is a promising technique for in~vivo medical imaging, offering lower radiation doses than traditional attenuation-based imaging. Previous studies have focused on X-ray energies below 50 keV for small-animal imaging and mammography. Here, we investigate the feasibility of high-energy propagation-based computed tomography for human adult-scale lung imaging at the Australian Synchrotron's Imaging and Medical Beamline. This facility is uniquely positioned for human lung imaging, offering a large field of view, high X-ray energies, and supporting clinical infrastructure. We imaged an anthropomorphic chest phantom (LungMan) between 50 keV and 80 keV across the range of possible sample-to-detector distances, with a photon-counting and an integrating detector. Strong phase-contrast fringes were observed with the photon-counting detector, even at high X-ray energies and a large pixel size relative to previous work, whereas the integrating detector with lower spatial resolution showed no clear phase effects. Measured X-ray phase-shifting properties of LungMan aligned well with reference soft tissue values, validating the phantom for phase-contrast studies. Imaging quality assessments suggest an optimal configuration at approximately 70 keV and the longest available propagation distance of 7.5 m, indicating potential benefit in positioning the patient in an upstream hutch. This study represents the first step towards clinical adult lung imaging at the Australian Synchrotron.
format Preprint
id arxiv_https___arxiv_org_abs_2505_14075
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle High-energy X-ray phase-contrast CT of an adult human chest phantom
Ahlers, Jannis N.
D'Amico, Lorenzo
Bast, Henriette
Costello, Lucy F.
Donnelley, Martin
Alloo, Samantha J.
Harker, Stephanie A.
How, Ying Ying
Croughan, Michelle K.
Pollock, James A.
Hausermann, Daniel
Maksimenko, Anton
Hall, Christopher
Gureyev, Timur E.
Nesterets, Yakov I.
Kitchen, Marcus J.
Pavlov, Konstantin M.
Morgan, Kaye S.
Medical Physics
Propagation-based phase-contrast X-ray imaging is a promising technique for in~vivo medical imaging, offering lower radiation doses than traditional attenuation-based imaging. Previous studies have focused on X-ray energies below 50 keV for small-animal imaging and mammography. Here, we investigate the feasibility of high-energy propagation-based computed tomography for human adult-scale lung imaging at the Australian Synchrotron's Imaging and Medical Beamline. This facility is uniquely positioned for human lung imaging, offering a large field of view, high X-ray energies, and supporting clinical infrastructure. We imaged an anthropomorphic chest phantom (LungMan) between 50 keV and 80 keV across the range of possible sample-to-detector distances, with a photon-counting and an integrating detector. Strong phase-contrast fringes were observed with the photon-counting detector, even at high X-ray energies and a large pixel size relative to previous work, whereas the integrating detector with lower spatial resolution showed no clear phase effects. Measured X-ray phase-shifting properties of LungMan aligned well with reference soft tissue values, validating the phantom for phase-contrast studies. Imaging quality assessments suggest an optimal configuration at approximately 70 keV and the longest available propagation distance of 7.5 m, indicating potential benefit in positioning the patient in an upstream hutch. This study represents the first step towards clinical adult lung imaging at the Australian Synchrotron.