Characterization of the Hamamatsu R12699-406-M4 Photomultiplier Tube in Cold Xenon Environments

Fuente: arXiv
Enregistré dans:
Détails bibliographiques
Auteurs principaux: Adrover, M., Baudis, L., Bismark, A., Colijn, A. P., Cuenca-García, J. J., Decowski, M. P., Flierman, M., Hollander, T. den
Format: Preprint
Publié: 2025
Sujets:
Accès en ligne:
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!
_version_ 1866915646093131776
author Adrover, M.
Baudis, L.
Bismark, A.
Colijn, A. P.
Cuenca-García, J. J.
Decowski, M. P.
Flierman, M.
Hollander, T. den
author_facet Adrover, M.
Baudis, L.
Bismark, A.
Colijn, A. P.
Cuenca-García, J. J.
Decowski, M. P.
Flierman, M.
Hollander, T. den
contents The Hamamatsu R12699-406-M2 is a $2\times2$ multi-anode 2-inch photomultiplier tube that offers a compact form factor, low intrinsic radioactivity, and high photocathode coverage. These characteristics make it a promising candidate for next-generation xenon-based direct detection dark matter experiments, such as XLZD and PandaX-xT. We present a detailed characterization of this photosensor operated in cold xenon environments, focusing on its single photoelectron response, dark count rate, light emission, and afterpulsing behavior. The device demonstrated a gain exceeding $2\cdot 10^6$ at the nominal voltage of -1.0 kV, along with a low dark count rate of $(0.4\pm0.2)\;\text{Hz/cm}^2$. Due to the compact design, afterpulses exhibited short delay times, resulting in some cases in an overlap with the light-induced signal. To evaluate its applicability in a realistic detector environment, two R12699-406-M2 units were deployed in a small-scale dual-phase xenon time projection chamber. The segmented $2\times2$ anode structure enabled lateral position reconstruction using a single photomultiplier tube, highlighting the potential of the sensor for effective event localization in future detectors.
format Preprint
id arxiv_https___arxiv_org_abs_2506_04844
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle Characterization of the Hamamatsu R12699-406-M4 Photomultiplier Tube in Cold Xenon Environments
Adrover, M.
Baudis, L.
Bismark, A.
Colijn, A. P.
Cuenca-García, J. J.
Decowski, M. P.
Flierman, M.
Hollander, T. den
Instrumentation and Detectors
Instrumentation and Methods for Astrophysics
The Hamamatsu R12699-406-M2 is a $2\times2$ multi-anode 2-inch photomultiplier tube that offers a compact form factor, low intrinsic radioactivity, and high photocathode coverage. These characteristics make it a promising candidate for next-generation xenon-based direct detection dark matter experiments, such as XLZD and PandaX-xT. We present a detailed characterization of this photosensor operated in cold xenon environments, focusing on its single photoelectron response, dark count rate, light emission, and afterpulsing behavior. The device demonstrated a gain exceeding $2\cdot 10^6$ at the nominal voltage of -1.0 kV, along with a low dark count rate of $(0.4\pm0.2)\;\text{Hz/cm}^2$. Due to the compact design, afterpulses exhibited short delay times, resulting in some cases in an overlap with the light-induced signal. To evaluate its applicability in a realistic detector environment, two R12699-406-M2 units were deployed in a small-scale dual-phase xenon time projection chamber. The segmented $2\times2$ anode structure enabled lateral position reconstruction using a single photomultiplier tube, highlighting the potential of the sensor for effective event localization in future detectors.
title Characterization of the Hamamatsu R12699-406-M4 Photomultiplier Tube in Cold Xenon Environments
topic Instrumentation and Detectors
Instrumentation and Methods for Astrophysics
url https://arxiv.org/abs/2506.04844