Leaky-wave Coil Element with Improved Tx-efficiency for 7 T MRI Using a Non-Uniform Current Design

Fuente: arXiv
Enregistré dans:
Détails bibliographiques
Auteurs principaux: Popova, K., Balafenidev, R., Svejda, J. T., Rennings, A., Raaijmakers, A. J., Collins, C. M., Lattanzi, R., Erni, D., Solomakha, G.
Format: Preprint
Publié: 2025
Sujets:
Accès en ligne:
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!
_version_ 1866914410712268800
author Popova, K.
Balafenidev, R.
Svejda, J. T.
Rennings, A.
Raaijmakers, A. J.
Collins, C. M.
Lattanzi, R.
Erni, D.
Solomakha, G.
author_facet Popova, K.
Balafenidev, R.
Svejda, J. T.
Rennings, A.
Raaijmakers, A. J.
Collins, C. M.
Lattanzi, R.
Erni, D.
Solomakha, G.
contents Imaging of the human body at ultra-high fields (static magnetic field B0>7 Tesla) is challenging due to the radio-frequency field inhomogeneities in the human body tissues caused by the short wavelength. These effects could be partially mitigated using an array of antennas and by parallel transmission allowing for control of the radio-frequency field distribution in the region of interest. All commonly-used radio-frequency arrays for ultra-high field MRI consist of resonant elements: dipoles, TEM-resonators, loops and individual slots. All these elements rely on standing wave excitation, in the sense that they are resonant devices that produce a field pattern with a constant phase distribution along the commensurable conductor elements. However, it was shown previously, that a non-uniform phase of surface current is required to reach the ultimate intrinsic signal-to-noise ratio or a maximized signal in the desired region of interest. In our work we propose to use a previously demonstrated non-resonant leaky-wave approach to control the phase of currents in radio-frequency coil conductors to increase the B1+ field in the center or in the region of interest. Using this approach, we developed a radio-frequency coil based on a leaky-wave antenna approach with optimized surface current distribution resulting in stronger B1+ in the desired region compared to e.g. a fractionated dipole.
format Preprint
id arxiv_https___arxiv_org_abs_2507_19286
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle Leaky-wave Coil Element with Improved Tx-efficiency for 7 T MRI Using a Non-Uniform Current Design
Popova, K.
Balafenidev, R.
Svejda, J. T.
Rennings, A.
Raaijmakers, A. J.
Collins, C. M.
Lattanzi, R.
Erni, D.
Solomakha, G.
Applied Physics
Imaging of the human body at ultra-high fields (static magnetic field B0>7 Tesla) is challenging due to the radio-frequency field inhomogeneities in the human body tissues caused by the short wavelength. These effects could be partially mitigated using an array of antennas and by parallel transmission allowing for control of the radio-frequency field distribution in the region of interest. All commonly-used radio-frequency arrays for ultra-high field MRI consist of resonant elements: dipoles, TEM-resonators, loops and individual slots. All these elements rely on standing wave excitation, in the sense that they are resonant devices that produce a field pattern with a constant phase distribution along the commensurable conductor elements. However, it was shown previously, that a non-uniform phase of surface current is required to reach the ultimate intrinsic signal-to-noise ratio or a maximized signal in the desired region of interest. In our work we propose to use a previously demonstrated non-resonant leaky-wave approach to control the phase of currents in radio-frequency coil conductors to increase the B1+ field in the center or in the region of interest. Using this approach, we developed a radio-frequency coil based on a leaky-wave antenna approach with optimized surface current distribution resulting in stronger B1+ in the desired region compared to e.g. a fractionated dipole.
title Leaky-wave Coil Element with Improved Tx-efficiency for 7 T MRI Using a Non-Uniform Current Design
topic Applied Physics
url https://arxiv.org/abs/2507.19286