Comparative study of YSO, GaGG, and BGO scintillators coupled to a SiPM array for gamma-ray spectroscopy

Fuente: arXiv
Enregistré dans:
Détails bibliographiques
Auteurs principaux: Nuruyev, S., Ahmadov, G., Berikov, D., Holik, M., Ahmadov, F., Sadigov, A., Akbarov, R., Naghiyev, J., Nuruyev, I., Mammadli, A.
Format: Preprint
Publié: 2025
Sujets:
Accès en ligne:
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!
_version_ 1866908568854200320
author Nuruyev, S.
Ahmadov, G.
Berikov, D.
Holik, M.
Ahmadov, F.
Sadigov, A.
Akbarov, R.
Naghiyev, J.
Nuruyev, I.
Mammadli, A.
author_facet Nuruyev, S.
Ahmadov, G.
Berikov, D.
Holik, M.
Ahmadov, F.
Sadigov, A.
Akbarov, R.
Naghiyev, J.
Nuruyev, I.
Mammadli, A.
contents A compact gamma-ray detector module was developed and characterized for spectroscopic applications across a wide energy range. The detector comprises a 4x4 array of silicon photomultipliers (SiPMs, Hamamatsu MPPC S13360-3050) coupled with three different inorganic scintillators (YSO, GaGG, BGO). The performance of the detector was evaluated over an energy range from 32 keV to 2500 keV using various standard gamma-ray point sources ({241}Am, {137}Cs, {22}Na, {60}Co, {228}Th). All three detector configurations demonstrated linearity between pulse height and gamma-ray energy over most of the tested range. GAGG exhibited the most consistent linear behavior up to 2500 keV. YSO with a faster decay constant showed linearity up to ~2000 keV, followed by slight saturation effects likely due to SiPM microcell occupancy limits at higher photon flux. BGO, characterized by a low light output and longer decay time, demonstrated stable linearity up to 2500 keV but reduced sensitivity below 300 keV. The measured energy resolution at 662 keV was 8.17% for GAGG, 9.3% for YSO, and 10.2% for BGO, consistent with their light yield and photon detection efficiency (PDE) - matching characteristics. The obtained results showed the trade-offs between scintillator materials in terms of energy resolution, dynamic range, and linearity, and confirm the suitability of the tested configurations for compact gamma-ray spectroscopy applications across the studied energy range.
format Preprint
id arxiv_https___arxiv_org_abs_2509_26252
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle Comparative study of YSO, GaGG, and BGO scintillators coupled to a SiPM array for gamma-ray spectroscopy
Nuruyev, S.
Ahmadov, G.
Berikov, D.
Holik, M.
Ahmadov, F.
Sadigov, A.
Akbarov, R.
Naghiyev, J.
Nuruyev, I.
Mammadli, A.
Instrumentation and Detectors
Applied Physics
A compact gamma-ray detector module was developed and characterized for spectroscopic applications across a wide energy range. The detector comprises a 4x4 array of silicon photomultipliers (SiPMs, Hamamatsu MPPC S13360-3050) coupled with three different inorganic scintillators (YSO, GaGG, BGO). The performance of the detector was evaluated over an energy range from 32 keV to 2500 keV using various standard gamma-ray point sources ({241}Am, {137}Cs, {22}Na, {60}Co, {228}Th). All three detector configurations demonstrated linearity between pulse height and gamma-ray energy over most of the tested range. GAGG exhibited the most consistent linear behavior up to 2500 keV. YSO with a faster decay constant showed linearity up to ~2000 keV, followed by slight saturation effects likely due to SiPM microcell occupancy limits at higher photon flux. BGO, characterized by a low light output and longer decay time, demonstrated stable linearity up to 2500 keV but reduced sensitivity below 300 keV. The measured energy resolution at 662 keV was 8.17% for GAGG, 9.3% for YSO, and 10.2% for BGO, consistent with their light yield and photon detection efficiency (PDE) - matching characteristics. The obtained results showed the trade-offs between scintillator materials in terms of energy resolution, dynamic range, and linearity, and confirm the suitability of the tested configurations for compact gamma-ray spectroscopy applications across the studied energy range.
title Comparative study of YSO, GaGG, and BGO scintillators coupled to a SiPM array for gamma-ray spectroscopy
topic Instrumentation and Detectors
Applied Physics
url https://arxiv.org/abs/2509.26252