Development of a Hermetic Gaseous Xenon Detector for Suppressing External Radon Background

Fuente: arXiv
Salvato in:
Dettagli Bibliografici
Autori principali: Miyata, Ryuta, Fujikawa, Koki, Harata, Rina, Itow, Yoshitaka, Kazama, Shingo, Kobayashi, Masatoshi
Natura: Preprint
Pubblicazione: 2025
Soggetti:
Accesso online:
Tags: Aggiungi Tag
Nessun Tag, puoi essere il primo ad aggiungerne!!
_version_ 1866908629951578112
author Miyata, Ryuta
Fujikawa, Koki
Harata, Rina
Itow, Yoshitaka
Kazama, Shingo
Kobayashi, Masatoshi
author_facet Miyata, Ryuta
Fujikawa, Koki
Harata, Rina
Itow, Yoshitaka
Kazama, Shingo
Kobayashi, Masatoshi
contents Radon-induced backgrounds, particularly from $^{222}$Rn and its beta-emitting progeny, present a critical challenge for next-generation liquid xenon (LXe) detectors aimed at probing dark matter down to the neutrino fog. To address this, we developed a compact hermetic gaseous xenon (GXe) detector. This device physically isolates the active volume from external radon sources by using a PTFE vessel sealed between two quartz flanges with mechanically compressed ePTFE gaskets. To quantify radon sealing performance, we implemented a dual-loop GXe circulation system and conducted a 670-hour radon-injection measurement campaign. Radon ingress into the hermetic detector was monitored using electrostatic radon detectors and photomultiplier tubes (PMTs). From these two independent measurements, the steady-state ratios of the radon concentrations inside the hermetic detector to those outside were estimated to be $ (1.1 \pm 0.1) \times 10^{-2} $ and $ (1.1 \pm 0.2) \times 10^{-2} $, corresponding to radon-leakage flows of $ (2.9 \pm 0.3) \times 10^{-11} $ and $ (2.6 \pm 0.4) \times 10^{-11} $ $ \thinspace $m$^{3}$ $ \thinspace $ $\mathrm{s}^{-1}$, respectively. An extrapolation to a 60-tonne LXe TPC such as XLZD suggests that the radon leakage could amount to $ 1.2 \times 10^{-2} $ $ \thinspace $mBq, which is negligible compared to the expected natural radon emanation inside the detector, typically 3$ \thinspace $mBq. These results demonstrate that flange-based mechanical sealing provides an effective solution for realizing radon-isolated inner detectors in large-scale LXe experiments.
format Preprint
id arxiv_https___arxiv_org_abs_2511_03250
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle Development of a Hermetic Gaseous Xenon Detector for Suppressing External Radon Background
Miyata, Ryuta
Fujikawa, Koki
Harata, Rina
Itow, Yoshitaka
Kazama, Shingo
Kobayashi, Masatoshi
Instrumentation and Detectors
High Energy Physics - Experiment
Radon-induced backgrounds, particularly from $^{222}$Rn and its beta-emitting progeny, present a critical challenge for next-generation liquid xenon (LXe) detectors aimed at probing dark matter down to the neutrino fog. To address this, we developed a compact hermetic gaseous xenon (GXe) detector. This device physically isolates the active volume from external radon sources by using a PTFE vessel sealed between two quartz flanges with mechanically compressed ePTFE gaskets. To quantify radon sealing performance, we implemented a dual-loop GXe circulation system and conducted a 670-hour radon-injection measurement campaign. Radon ingress into the hermetic detector was monitored using electrostatic radon detectors and photomultiplier tubes (PMTs). From these two independent measurements, the steady-state ratios of the radon concentrations inside the hermetic detector to those outside were estimated to be $ (1.1 \pm 0.1) \times 10^{-2} $ and $ (1.1 \pm 0.2) \times 10^{-2} $, corresponding to radon-leakage flows of $ (2.9 \pm 0.3) \times 10^{-11} $ and $ (2.6 \pm 0.4) \times 10^{-11} $ $ \thinspace $m$^{3}$ $ \thinspace $ $\mathrm{s}^{-1}$, respectively. An extrapolation to a 60-tonne LXe TPC such as XLZD suggests that the radon leakage could amount to $ 1.2 \times 10^{-2} $ $ \thinspace $mBq, which is negligible compared to the expected natural radon emanation inside the detector, typically 3$ \thinspace $mBq. These results demonstrate that flange-based mechanical sealing provides an effective solution for realizing radon-isolated inner detectors in large-scale LXe experiments.
title Development of a Hermetic Gaseous Xenon Detector for Suppressing External Radon Background
topic Instrumentation and Detectors
High Energy Physics - Experiment
url https://arxiv.org/abs/2511.03250