Performance Evaluation of a Position-Sensitive SiPM-based Gamma Camera for Intraoperative Imaging

Fuente: arXiv
Saved in:
Bibliographic Details
Main Authors: Raiola, Aramis, Acerbi, Fabio, Alispach, Cyril, della Volpe, Domenico, Arabi, Hossein, Gola, Alberto, Zaidi, Habib
Format: Preprint
Published: 2025
Subjects:
Online Access:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1866912863541526528
author Raiola, Aramis
Acerbi, Fabio
Alispach, Cyril
della Volpe, Domenico
Arabi, Hossein
Gola, Alberto
Zaidi, Habib
author_facet Raiola, Aramis
Acerbi, Fabio
Alispach, Cyril
della Volpe, Domenico
Arabi, Hossein
Gola, Alberto
Zaidi, Habib
contents The POSiCS camera is a handheld, small field-of-view gamma camera developed for multipurpose use in radio-guided surgery (RGS), with sentinel lymph node biopsy (SLNB) as its benchmark application. This compact and lightweight detector (weighing approximately 350 g) can map tissues labeled with Tc-99m nanocolloids and guide surgeons to the location of target lesions. By enabling intraoperative visualization in close proximity to the surgical field, its primary objective is to minimize surgical interventional invasiveness and operative time, thereby enhancing localization accuracy and reducing the incidence of post-operative complications. The design and components of the POSiCS camera emphasize ergonomic handling and compactness, providing, at the same time, rapid image formation and a spatial resolution of a few millimeters. These features are compatible with routine operating-room workflow, including wireless communication with the computer and a real-time display to support surgeon decision-making. The spatial resolution measured at a source-detector distance of 0 cm was 1.9 +/- 0.1 mm for the high-sensitivity mode and 1.4 +/- 0.1 mm for the high-resolution mode. The system sensitivity at 2 cm was evaluated as 481 +/- 14 cps/MBq (high sensitivity) and 134 +/- 8 cps/MBq (high resolution). For both working modes, we report an energy resolution of approximately 20 percent, even though the high-resolution collimator exhibits an increased scattered component due to the larger amount of tungsten.
format Preprint
id arxiv_https___arxiv_org_abs_2511_03493
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle Performance Evaluation of a Position-Sensitive SiPM-based Gamma Camera for Intraoperative Imaging
Raiola, Aramis
Acerbi, Fabio
Alispach, Cyril
della Volpe, Domenico
Arabi, Hossein
Gola, Alberto
Zaidi, Habib
Medical Physics
The POSiCS camera is a handheld, small field-of-view gamma camera developed for multipurpose use in radio-guided surgery (RGS), with sentinel lymph node biopsy (SLNB) as its benchmark application. This compact and lightweight detector (weighing approximately 350 g) can map tissues labeled with Tc-99m nanocolloids and guide surgeons to the location of target lesions. By enabling intraoperative visualization in close proximity to the surgical field, its primary objective is to minimize surgical interventional invasiveness and operative time, thereby enhancing localization accuracy and reducing the incidence of post-operative complications. The design and components of the POSiCS camera emphasize ergonomic handling and compactness, providing, at the same time, rapid image formation and a spatial resolution of a few millimeters. These features are compatible with routine operating-room workflow, including wireless communication with the computer and a real-time display to support surgeon decision-making. The spatial resolution measured at a source-detector distance of 0 cm was 1.9 +/- 0.1 mm for the high-sensitivity mode and 1.4 +/- 0.1 mm for the high-resolution mode. The system sensitivity at 2 cm was evaluated as 481 +/- 14 cps/MBq (high sensitivity) and 134 +/- 8 cps/MBq (high resolution). For both working modes, we report an energy resolution of approximately 20 percent, even though the high-resolution collimator exhibits an increased scattered component due to the larger amount of tungsten.
title Performance Evaluation of a Position-Sensitive SiPM-based Gamma Camera for Intraoperative Imaging
topic Medical Physics
url https://arxiv.org/abs/2511.03493