Lightweight Deep Autoencoder for ECG Denoising with Morphology Preservation and Near Real-Time Hardware Deployment

Fuente: arXiv
Saved in:
Bibliographic Details
Main Authors: Nejad, Mahdi Pirayesh Shirazi, Hicks, David, Valentine, Matt, Chon, Ki H.
Format: Preprint
Published: 2025
Subjects:
Online Access:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1866915620968202240
author Nejad, Mahdi Pirayesh Shirazi
Hicks, David
Valentine, Matt
Chon, Ki H.
author_facet Nejad, Mahdi Pirayesh Shirazi
Hicks, David
Valentine, Matt
Chon, Ki H.
contents Electrocardiogram (ECG) signals are often degraded by various noise sources such as baseline wander, motion artifacts, and electromyographic interference, posing a major challenge in clinical settings. This paper presents a lightweight deep learning-based denoising framework, forming a compact autoencoder architecture. The model was trained under severe noise conditions (-5 dB signal-to-noise ratio (SNR)) using a rigorously partitioned dataset to ensure no data leakage and robust generalization. Extensive evaluations were conducted across seven noise configurations and three SNR levels (-5 dB, 0 dB, and +5 dB), showing consistent denoising performance with minimal morphological distortion, critical for maintaining diagnostic integrity. In particular, tests on clinically vital rhythms such as ventricular tachycardia (VT) and ventricular fibrillation (VF) confirm that the proposed model effectively suppresses noise without altering arrhythmic features essential for diagnosis. Visual and quantitative assessments, including SNR improvement, RMSE, and correlation metrics, validate the model's efficacy in preserving waveform fidelity. To demonstrate real-world applicability, the model was deployed on a Raspberry Pi 4 using TensorFlow Lite with float16 precision. Inference latency was measured at just 1.41 seconds per 14-second ECG segment, indicating feasibility for near-real-time use in edge devices. Overall, this study introduces a lightweight, hardware-validated, and morphologically reliable ECG denoising solution suitable for integration into portable or wearable healthcare systems.
format Preprint
id arxiv_https___arxiv_org_abs_2511_12478
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle Lightweight Deep Autoencoder for ECG Denoising with Morphology Preservation and Near Real-Time Hardware Deployment
Nejad, Mahdi Pirayesh Shirazi
Hicks, David
Valentine, Matt
Chon, Ki H.
Signal Processing
Electrocardiogram (ECG) signals are often degraded by various noise sources such as baseline wander, motion artifacts, and electromyographic interference, posing a major challenge in clinical settings. This paper presents a lightweight deep learning-based denoising framework, forming a compact autoencoder architecture. The model was trained under severe noise conditions (-5 dB signal-to-noise ratio (SNR)) using a rigorously partitioned dataset to ensure no data leakage and robust generalization. Extensive evaluations were conducted across seven noise configurations and three SNR levels (-5 dB, 0 dB, and +5 dB), showing consistent denoising performance with minimal morphological distortion, critical for maintaining diagnostic integrity. In particular, tests on clinically vital rhythms such as ventricular tachycardia (VT) and ventricular fibrillation (VF) confirm that the proposed model effectively suppresses noise without altering arrhythmic features essential for diagnosis. Visual and quantitative assessments, including SNR improvement, RMSE, and correlation metrics, validate the model's efficacy in preserving waveform fidelity. To demonstrate real-world applicability, the model was deployed on a Raspberry Pi 4 using TensorFlow Lite with float16 precision. Inference latency was measured at just 1.41 seconds per 14-second ECG segment, indicating feasibility for near-real-time use in edge devices. Overall, this study introduces a lightweight, hardware-validated, and morphologically reliable ECG denoising solution suitable for integration into portable or wearable healthcare systems.
title Lightweight Deep Autoencoder for ECG Denoising with Morphology Preservation and Near Real-Time Hardware Deployment
topic Signal Processing
url https://arxiv.org/abs/2511.12478