WFC3/UVIS EPER CTE 2009-2025

Fuente: arXiv
Saved in:
Bibliographic Details
Main Authors: O'Connor, Anne, Khandrika, Harish
Format: Preprint
Published: 2025
Subjects:
Online Access:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1866911277430865920
author O'Connor, Anne
Khandrika, Harish
author_facet O'Connor, Anne
Khandrika, Harish
contents In this report, we examine the behavior of Charge Transfer Efficiency (CTE) on the WFC3/UVIS detector over time as computed by the Extended Pixel Edge Response (EPER) technique, using internal calibration data acquired from 2009 through 2025. We find that the CTE has continued to decline as expected, with a steeper loss rate for lower signal levels. The lowest signal level (160e-) has continued to decline at a rate of 0.0001 per year, with a total overall decline of 0.0015. Analyses from 2016 and 2020 found that the rate of decline was not well fit by a linear function. This report verifies the rate of decline is instead better fit by a quadratic function (which results in the smallest min. and max. residuals, on average) or a cubic function (which has the best "goodness of fit" $χ^2$ and $R^2$ values). We continue to see periodic oscillations of the residuals around all three fit lines (linear, quadratic, and cubic) on which we perform a Lomb-Scargle periodogram analysis of the residuals. We find a periodicity of about 8 years for the residuals around the linear fit lines and about 9 years for the quadratic and cubic fit lines.
format Preprint
id arxiv_https___arxiv_org_abs_2511_16391
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle WFC3/UVIS EPER CTE 2009-2025
O'Connor, Anne
Khandrika, Harish
Instrumentation and Methods for Astrophysics
In this report, we examine the behavior of Charge Transfer Efficiency (CTE) on the WFC3/UVIS detector over time as computed by the Extended Pixel Edge Response (EPER) technique, using internal calibration data acquired from 2009 through 2025. We find that the CTE has continued to decline as expected, with a steeper loss rate for lower signal levels. The lowest signal level (160e-) has continued to decline at a rate of 0.0001 per year, with a total overall decline of 0.0015. Analyses from 2016 and 2020 found that the rate of decline was not well fit by a linear function. This report verifies the rate of decline is instead better fit by a quadratic function (which results in the smallest min. and max. residuals, on average) or a cubic function (which has the best "goodness of fit" $χ^2$ and $R^2$ values). We continue to see periodic oscillations of the residuals around all three fit lines (linear, quadratic, and cubic) on which we perform a Lomb-Scargle periodogram analysis of the residuals. We find a periodicity of about 8 years for the residuals around the linear fit lines and about 9 years for the quadratic and cubic fit lines.
title WFC3/UVIS EPER CTE 2009-2025
topic Instrumentation and Methods for Astrophysics
url https://arxiv.org/abs/2511.16391