FENCE: Flexible Electric Noise reduCtion Endo-shield for the Suppression of Electromagnetic Interference in Low-Field MRI

Fuente: arXiv
Saved in:
Bibliographic Details
Main Authors: Pfitzer, Julia, Uecker, Martin, Scharfetter, Hermann
Format: Preprint
Published: 2025
Subjects:
Online Access:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1866911712173621248
author Pfitzer, Julia
Uecker, Martin
Scharfetter, Hermann
author_facet Pfitzer, Julia
Uecker, Martin
Scharfetter, Hermann
contents Electromagnetic interference (EMI) is a significant challenge for low-field MRI systems operating without conventional Faraday-shielded rooms. Traditional EMI mitigation approaches include external shields, subject grounding via electrodes, or active noise cancellation requiring synchronized receive channels. These methods either limit portability, introduce patient discomfort, or demand advanced hardware. In this work, we start from the hypothesis that EMI primarily couples capacitively from the body to the RF coil. We investigated two methods of blocking capacitive coupling while preserving inductive MRI signal detection: First, we employed capacitive segmentation of the RF coil and studied its effect on EMI coupling. Second, we present FENCE (Flexible Electromagnetic Noise reduCtion Endo-shield), a novel approach blocking capacitive coupling using flexible PCB shields placed inside the RF coil. FENCE can be retrofitted to existing RF coils. Finite element (FE) simulations were used to estimate the expected shielding performance and the impact on RF coil losses prior to practical implementation. Testing in various realistic scenarios then demonstrated that the combination of FENCE with segmented solenoid coils is effective against both environmental noise sources and controlled EMI. In phantom experiments, FENCE significantly improved imaging performance and reduced EMI levels to near-baseline levels with 9% reduction in coil quality factor (Q factor), showing good agreement with the predictions from the FE simulations. In-vivo head imaging confirmed these results across diverse electromagnetic environments significantly improving imaging performance while showing an ~18% decrease in Q factor. FENCE provides a simple method for EMI mitigation in low-field MRI, enhancing image quality while maintaining system portability and accessibility.
format Preprint
id arxiv_https___arxiv_org_abs_2512_04889
institution arXiv
publishDate 2025
record_format arxiv
spellingShingle FENCE: Flexible Electric Noise reduCtion Endo-shield for the Suppression of Electromagnetic Interference in Low-Field MRI
Pfitzer, Julia
Uecker, Martin
Scharfetter, Hermann
Medical Physics
Electromagnetic interference (EMI) is a significant challenge for low-field MRI systems operating without conventional Faraday-shielded rooms. Traditional EMI mitigation approaches include external shields, subject grounding via electrodes, or active noise cancellation requiring synchronized receive channels. These methods either limit portability, introduce patient discomfort, or demand advanced hardware. In this work, we start from the hypothesis that EMI primarily couples capacitively from the body to the RF coil. We investigated two methods of blocking capacitive coupling while preserving inductive MRI signal detection: First, we employed capacitive segmentation of the RF coil and studied its effect on EMI coupling. Second, we present FENCE (Flexible Electromagnetic Noise reduCtion Endo-shield), a novel approach blocking capacitive coupling using flexible PCB shields placed inside the RF coil. FENCE can be retrofitted to existing RF coils. Finite element (FE) simulations were used to estimate the expected shielding performance and the impact on RF coil losses prior to practical implementation. Testing in various realistic scenarios then demonstrated that the combination of FENCE with segmented solenoid coils is effective against both environmental noise sources and controlled EMI. In phantom experiments, FENCE significantly improved imaging performance and reduced EMI levels to near-baseline levels with 9% reduction in coil quality factor (Q factor), showing good agreement with the predictions from the FE simulations. In-vivo head imaging confirmed these results across diverse electromagnetic environments significantly improving imaging performance while showing an ~18% decrease in Q factor. FENCE provides a simple method for EMI mitigation in low-field MRI, enhancing image quality while maintaining system portability and accessibility.
title FENCE: Flexible Electric Noise reduCtion Endo-shield for the Suppression of Electromagnetic Interference in Low-Field MRI
topic Medical Physics
url https://arxiv.org/abs/2512.04889