Hemodynamic Performance and Blood Damage of the Intra-Aortic Pumps: A CFD-Based Investigation

Fuente: arXiv
Salvato in:
Dettagli Bibliografici
Autori principali: Aycan, Osman, Park, Yeojin, Kadem, Lyes
Natura: Preprint
Pubblicazione: 2026
Soggetti:
Accesso online:
Tags: Aggiungi Tag
Nessun Tag, puoi essere il primo ad aggiungerne!!
_version_ 1866917364863336448
author Aycan, Osman
Park, Yeojin
Kadem, Lyes
author_facet Aycan, Osman
Park, Yeojin
Kadem, Lyes
contents Three intra-aortic blood pumps were evaluated and compared using CFD simulations. The aim of this study is to evaluate the hemodynamic performance and calculate the hemolytic potential of the pumps. The flow fields generated by the pumps were simulated using CFD with a wall-modeled large eddy simulation (WMLES) approach. The simulations produced pressure-flow curves, hydraulic efficiency, shear stress distributions, and hemolysis predictions. A grid study was conducted using the Grid Convergence Index (GCI) method, and a new dimensionless parameter, the Hemolytic Number (HN), was introduced as a standardized metric to compare hemolysis and universal pump performance. The impeller-driven pump had the highest-pressure head (~800 Pa at 4 L/min) and the best hydraulic efficiency (~6% at 14 L/min), outperforming both the single (maximum 2.7%) and triplet (maximum 2.2%) pumps. The NIH values were also lowest for the impeller pump (NIH = 0.0035 g/100L), indicating high hemocompatibility. Both the single and triplet pumps showed regions of recirculation, particularly at lower flow rates. A smaller HN indicates better hemocompatibility; for the impeller-driven pump, HN remains below 1 across the investigated flow rates. Overall, the impeller-driven pump outperformed the other designs in terms of both hemodynamic and hemolytic measurements. The findings provide valuable insights for the future development of intra-aortic pumps and the personalized selection of devices for individual patients.
format Preprint
id arxiv_https___arxiv_org_abs_2603_26911
institution arXiv
publishDate 2026
record_format arxiv
spellingShingle Hemodynamic Performance and Blood Damage of the Intra-Aortic Pumps: A CFD-Based Investigation
Aycan, Osman
Park, Yeojin
Kadem, Lyes
Medical Physics
Fluid Dynamics
Three intra-aortic blood pumps were evaluated and compared using CFD simulations. The aim of this study is to evaluate the hemodynamic performance and calculate the hemolytic potential of the pumps. The flow fields generated by the pumps were simulated using CFD with a wall-modeled large eddy simulation (WMLES) approach. The simulations produced pressure-flow curves, hydraulic efficiency, shear stress distributions, and hemolysis predictions. A grid study was conducted using the Grid Convergence Index (GCI) method, and a new dimensionless parameter, the Hemolytic Number (HN), was introduced as a standardized metric to compare hemolysis and universal pump performance. The impeller-driven pump had the highest-pressure head (~800 Pa at 4 L/min) and the best hydraulic efficiency (~6% at 14 L/min), outperforming both the single (maximum 2.7%) and triplet (maximum 2.2%) pumps. The NIH values were also lowest for the impeller pump (NIH = 0.0035 g/100L), indicating high hemocompatibility. Both the single and triplet pumps showed regions of recirculation, particularly at lower flow rates. A smaller HN indicates better hemocompatibility; for the impeller-driven pump, HN remains below 1 across the investigated flow rates. Overall, the impeller-driven pump outperformed the other designs in terms of both hemodynamic and hemolytic measurements. The findings provide valuable insights for the future development of intra-aortic pumps and the personalized selection of devices for individual patients.