Design and construction of a gallium fixed-point blackbody at CENAM

Fuente: Redalyc
Enregistré dans:
Détails bibliographiques
Auteur principal: D. Cárdenas-García
Format: Artículo científico
Langue:en
Publié: Sociedad Mexicana de Física A.C. 2015
Sujets:
Accès en ligne:
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!
_version_ 1876486224785440768
author D. Cárdenas-García
author_facet D. Cárdenas-García
contents Design and construction of a gallium fixed-point blackbody at CENAM D. Cárdenas-García Física, Astronomía y Matemáticas fixed point blackbody radiation thermometry International temperature scale of 1990 For temperatures below silver fixed-point defined by the International Temperature Scale of 1990, a transfer radiation thermometer can be calibrated using either of two calibration schemes: a variable temperature blackbody with a standard platinum resistance thermometer as a reference, or with a set of fixed-point blackbodies. CENAM is presently working with the first scheme, and it is developing fixed-point blackbodies to have the capability to work with the second scheme too. For this purpose a gallium fixed-point blackbody to calibrate CENAM’s transfer radiation thermometer was designed and constructed. The blackbody cavity has a cylindro-cone shape with effective emissivity equal to 0:9992§0:0004 in the 8 ¹m to 14 ¹m wavelength range. The radiance temperature of the gallium fixed-point blackbody was estimated to have an expanded uncertainty of 54 mK, with a coverage factor k = 2. 2015 artículo científico 0035-001X https://www.redalyc.org/articulo.oa?id=57036862001 en http://www.redalyc.org/revista.oa?id=570 Revista Mexicana de Física application/pdf Sociedad Mexicana de Física A.C. Revista Mexicana de Física (México) Num.2 Vol.61
format Artículo científico
id redalyc_57036862001
institution Redalyc
language en
publishDate 2015
publisher Sociedad Mexicana de Física A.C.
spellingShingle Design and construction of a gallium fixed-point blackbody at CENAM
D. Cárdenas-García
Física, Astronomía y Matemáticas
fixed
point blackbody
radiation thermometry
International temperature scale of 1990
Design and construction of a gallium fixed-point blackbody at CENAM D. Cárdenas-García Física, Astronomía y Matemáticas fixed point blackbody radiation thermometry International temperature scale of 1990 For temperatures below silver fixed-point defined by the International Temperature Scale of 1990, a transfer radiation thermometer can be calibrated using either of two calibration schemes: a variable temperature blackbody with a standard platinum resistance thermometer as a reference, or with a set of fixed-point blackbodies. CENAM is presently working with the first scheme, and it is developing fixed-point blackbodies to have the capability to work with the second scheme too. For this purpose a gallium fixed-point blackbody to calibrate CENAM’s transfer radiation thermometer was designed and constructed. The blackbody cavity has a cylindro-cone shape with effective emissivity equal to 0:9992§0:0004 in the 8 ¹m to 14 ¹m wavelength range. The radiance temperature of the gallium fixed-point blackbody was estimated to have an expanded uncertainty of 54 mK, with a coverage factor k = 2. 2015 artículo científico 0035-001X https://www.redalyc.org/articulo.oa?id=57036862001 en http://www.redalyc.org/revista.oa?id=570 Revista Mexicana de Física application/pdf Sociedad Mexicana de Física A.C. Revista Mexicana de Física (México) Num.2 Vol.61
title Design and construction of a gallium fixed-point blackbody at CENAM
topic Física, Astronomía y Matemáticas
fixed
point blackbody
radiation thermometry
International temperature scale of 1990
url https://www.redalyc.org/articulo.oa?id=57036862001