Performance evaluation of a small‐animal PET scanner with 213 mm axis using NEMA NU 4‐2008

Fuente: Wiley Open Access
Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Qing Wei, Yingjie Wang, Xianchao Huang, Daowu Li, Yushuang Zheng, Peilin Wang, Xiaoli Sun, Pei Chai, Xiaorou Han, Shuangquan Liu, Baotong Feng, Wei Zhou, Xiangtao Zeng, Meiling Zhu, Zhiming Zhang, Long Wei
Format: Artículo Open Access
Veröffentlicht: Wiley 2024
Schlagworte:
Online-Zugang:
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
_version_ 1867011809545814016
author Qing Wei
Yingjie Wang
Xianchao Huang
Daowu Li
Yushuang Zheng
Peilin Wang
Xiaoli Sun
Pei Chai
Xiaorou Han
Shuangquan Liu
Baotong Feng
Wei Zhou
Xiangtao Zeng
Meiling Zhu
Zhiming Zhang
Long Wei
author_facet Qing Wei
Yingjie Wang
Xianchao Huang
Daowu Li
Yushuang Zheng
Peilin Wang
Xiaoli Sun
Pei Chai
Xiaorou Han
Shuangquan Liu
Baotong Feng
Wei Zhou
Xiangtao Zeng
Meiling Zhu
Zhiming Zhang
Long Wei
Qing Wei
Yingjie Wang
Xianchao Huang
Daowu Li
Yushuang Zheng
Peilin Wang
Xiaoli Sun
Pei Chai
Xiaorou Han
Shuangquan Liu
Baotong Feng
Wei Zhou
Xiangtao Zeng
Meiling Zhu
Zhiming Zhang
Long Wei
collection Wiley Open Access
contents Performance evaluation of a small‐animal PET scanner with 213 mm axis using NEMA NU 4‐2008 Qing Wei Yingjie Wang Xianchao Huang Daowu Li Yushuang Zheng Peilin Wang Xiaoli Sun Pei Chai Xiaorou Han Shuangquan Liu Baotong Feng Wei Zhou Xiangtao Zeng Meiling Zhu Zhiming Zhang Long Wei Medical Physics AbstractBackgroundLong‐axis positron emission tomography (PET) has emerged as one of the recent research directions in PET due to its ability to significantly enhance sensitivity and counting performance for low‐dose imaging, rapid imaging, and whole‐body dynamic imaging.PurposeThe PET system presented in this study is a long‐axis animal PET based on lutetium–yttrium orthosilicate and silicon photomultiplier, designed for whole‐body imaging in rats. It features a diameter of 143 mm and an axial length of 213.3 mm. This study evaluated the performance of this PET system in accordance with the National Electrical Manufacturers Association (NEMA) NU 4‐2008 standards.MethodsThe performance evaluation was conducted according to the NEMA NU 4‐2008 standards in terms of spatial resolution, sensitivity, counting rate performance, scatter fraction (SF) and image quality. In addition, a rat imaging study was conducted to assess the imaging capability of this PET system.ResultsThe average energy resolution of the PET system was 12.87%, the average coincidence timing resolution was 751 ps. The FWHM of spatial resolution reconstructed by filtered back projection and 3D‐OSEM‐PSF algorithm at 5 mm radial offset from the axial center were 1.65 and 0.88 mm. The peak absolute sensitivity measured by a point source at the center of the field of view was evaluated as 6.71% (361–661 keV) and 10.31% (250–750 keV). For the mouse‐like phantom, the SF was 11.0% and the peak noise equivalent counting rate (NECR) was 1193 kcps at 94.2 MBq (2.54 mCi). For the rat‐like phantom, the SF was 26.8% and the NECR was 682.5 kcps at 78.6 MBq (2.12 mCi).ConclusionsThe performance measurement results demonstrate that this PET system exhibits high sensitivity and count rate performance, making it potential for high‐quality whole‐body dynamic imaging of rats. 10.1002/mp.17469 http://onlinelibrary.wiley.com/termsAndConditions#vor
doi_str_mv 10.1002/mp.17469
format Artículo Open Access
id wiley_oa_10_1002_mp_17469
institution Wiley Open Access
license_str_mv http://onlinelibrary.wiley.com/termsAndConditions#vor
publishDate 2024
publisher Wiley
record_format wiley_oa
spellingShingle Performance evaluation of a small‐animal PET scanner with 213 mm axis using NEMA NU 4‐2008
Qing Wei
Yingjie Wang
Xianchao Huang
Daowu Li
Yushuang Zheng
Peilin Wang
Xiaoli Sun
Pei Chai
Xiaorou Han
Shuangquan Liu
Baotong Feng
Wei Zhou
Xiangtao Zeng
Meiling Zhu
Zhiming Zhang
Long Wei
Medical Physics
Performance evaluation of a small‐animal PET scanner with 213 mm axis using NEMA NU 4‐2008 Qing Wei Yingjie Wang Xianchao Huang Daowu Li Yushuang Zheng Peilin Wang Xiaoli Sun Pei Chai Xiaorou Han Shuangquan Liu Baotong Feng Wei Zhou Xiangtao Zeng Meiling Zhu Zhiming Zhang Long Wei Medical Physics AbstractBackgroundLong‐axis positron emission tomography (PET) has emerged as one of the recent research directions in PET due to its ability to significantly enhance sensitivity and counting performance for low‐dose imaging, rapid imaging, and whole‐body dynamic imaging.PurposeThe PET system presented in this study is a long‐axis animal PET based on lutetium–yttrium orthosilicate and silicon photomultiplier, designed for whole‐body imaging in rats. It features a diameter of 143 mm and an axial length of 213.3 mm. This study evaluated the performance of this PET system in accordance with the National Electrical Manufacturers Association (NEMA) NU 4‐2008 standards.MethodsThe performance evaluation was conducted according to the NEMA NU 4‐2008 standards in terms of spatial resolution, sensitivity, counting rate performance, scatter fraction (SF) and image quality. In addition, a rat imaging study was conducted to assess the imaging capability of this PET system.ResultsThe average energy resolution of the PET system was 12.87%, the average coincidence timing resolution was 751 ps. The FWHM of spatial resolution reconstructed by filtered back projection and 3D‐OSEM‐PSF algorithm at 5 mm radial offset from the axial center were 1.65 and 0.88 mm. The peak absolute sensitivity measured by a point source at the center of the field of view was evaluated as 6.71% (361–661 keV) and 10.31% (250–750 keV). For the mouse‐like phantom, the SF was 11.0% and the peak noise equivalent counting rate (NECR) was 1193 kcps at 94.2 MBq (2.54 mCi). For the rat‐like phantom, the SF was 26.8% and the NECR was 682.5 kcps at 78.6 MBq (2.12 mCi).ConclusionsThe performance measurement results demonstrate that this PET system exhibits high sensitivity and count rate performance, making it potential for high‐quality whole‐body dynamic imaging of rats. 10.1002/mp.17469 http://onlinelibrary.wiley.com/termsAndConditions#vor
title Performance evaluation of a small‐animal PET scanner with 213 mm axis using NEMA NU 4‐2008
topic Medical Physics