Development and validation of RP-HPLC method for determination of propofol in bulk and pharmaceutical dosage form.

Fuente: Zenodo
Guardado en:
Detalles Bibliográficos
Autores principales: Dr. Shweta Singh, Dr. Avadhesh Diwakar
Formato: Recurso digital
Publicado: Zenodo 2026
Acceso en línea:
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
_version_ 1866901251249143808
author Dr. Shweta Singh
Dr. Avadhesh Diwakar
author_facet Dr. Shweta Singh
Dr. Avadhesh Diwakar
contents <p><span>The aim of this study was to develop a simple, precise, accurate,<br>economic and reproducible RP-HPLC method for determination<br>of propofol in bulk and pharmaceutical dosage form and<br>validate as per ICH guidelines. In this analysis using C</span><span>18<br></span><span>column and its dimensions is 250 x 4.6 mm, 5µm with methanol<br>and water for injection as mobile phase in the ratio of 84:16 at<br>flow rate of 0.8 ml/min and detection was carried out at 275 nm<br>by using UV visible detector. The injection volume is 20 µl and<br>retention time was found to be 7min with 18 min as run time. In<br>HPLC method linearity was found in 10-50 µg/ml with<br>correlation coefficient 0.998. The percentage recovery was<br>found to be 99.1 to 100.3 %. The proposed method gave good</span> </p> <p><span>separation for propofol and method was validated for accuracy,<br>precision, linearity, limit of detection, limit of quantitation,<br>robustness and ruggedness. The result of study showed that the<br>proposed method RP-HPLC is suitable for routine analysis of<br>itraconazole in bulk and pharmaceutical dosage form.</span> </p>
format Recurso digital
id zenodo_https___doi_org_10_5281_zenodo_19149438
institution Zenodo
language
publishDate 2026
publisher Zenodo
record_format zenodo
spellingShingle Development and validation of RP-HPLC method for determination of propofol in bulk and pharmaceutical dosage form.
Dr. Shweta Singh
Dr. Avadhesh Diwakar
<p><span>The aim of this study was to develop a simple, precise, accurate,<br>economic and reproducible RP-HPLC method for determination<br>of propofol in bulk and pharmaceutical dosage form and<br>validate as per ICH guidelines. In this analysis using C</span><span>18<br></span><span>column and its dimensions is 250 x 4.6 mm, 5µm with methanol<br>and water for injection as mobile phase in the ratio of 84:16 at<br>flow rate of 0.8 ml/min and detection was carried out at 275 nm<br>by using UV visible detector. The injection volume is 20 µl and<br>retention time was found to be 7min with 18 min as run time. In<br>HPLC method linearity was found in 10-50 µg/ml with<br>correlation coefficient 0.998. The percentage recovery was<br>found to be 99.1 to 100.3 %. The proposed method gave good</span> </p> <p><span>separation for propofol and method was validated for accuracy,<br>precision, linearity, limit of detection, limit of quantitation,<br>robustness and ruggedness. The result of study showed that the<br>proposed method RP-HPLC is suitable for routine analysis of<br>itraconazole in bulk and pharmaceutical dosage form.</span> </p>
title Development and validation of RP-HPLC method for determination of propofol in bulk and pharmaceutical dosage form.
url https://doi.org/10.5281/zenodo.19149438